The 50 Turn 5500A/COIL is used as a tool for calibrating clamp type current meters that operate by two different principles ? as current transformers (AC only), and by the Hall Effect (Both AC and DC). It is impractical to calibrate 1000A rated current clamp meters using a 1000A source. However, by using the 50 turns of the 5500A/COIL in conjunction with a current calibrator you can effectively multiply the current of the current calibrator by a factor of 50 to support the calibration and verification of these clamp-on current meters.
型号名称 描述
5500A/COIL 50 匝电流线圈
本公司长期高价收购/回收:Agilent、HP、Anritsu、Advantest、R/S、/MARCONI网络分析仪、手机综合测试仪、无线电综合测试仪、万用表校准仪、频谱分析仪、视频分析仪、音频分析仪、噪声系数分析仪、蓝牙测试仪、视频图象信号源、LCR测试仪(电桥)、高低频信号源、示波器、万用表、手机程控电源、电子负载、频率计、功率计、仪器配件,GPIB卡、GPIB转USB线等等。欢迎来电咨询!!!
Supports the calibration of clamp-type current meters
The 50 Turn 5500A/COIL is used as a tool for calibrating clamp type current meters that operate by two different principles ? as current transformers (AC only), and by the Hall Effect (Both AC and DC). It is impractical to calibrate 1000A rated current clamp meters using a 1000A source. However, by using the 50 turns of the 5500A/COIL in conjunction with a current calibrator you can effectively multiply the current of the current calibrator by a factor of 50 to support the calibration and verification of these clamp-on current meters
General Specifications
Number of Turns
50
Maximum Current
11A rms, continuous
20A rms, 2 minutes
Maximum Duty Cycle Derating
< 11A, continuous
> 11A, 2 minutes ON, 8 minutes OFF
Maximum Voltage
3V rms
Frequency of Operation
DC, 45-440 Hz
(should not exceed rms Voltage rating)
For the 5500A Current Output:
DC, 45-65 Hz, 0-11A
65-440 Hz, 0-2.19999A
For the 5520A Current Output:
DC, 45-65 Hz, 0-20A
65-440 Hz, 0-2.99999A
This specification assumes that a Fluke 80i-600 Clamp Meter or equivalent is attached. Other clamp meters may limit the 5500A/5520A Current Output drive capability to less than the above.
Uncertainty Due to Clamp Meter/Coil Interaction
± (0.25% of effective output + 0.5A), for toroidal-wound current clamps, such as the Fluke 80I and 80I-1000.
± (0.50% of effective output + 0.5A), for current clamps like the Fluke 80i-kw, 80i-400, 80i-410, 80i-500, 80i-1010, Fluke 31, Fluke 33, or equivalent.
Minimum Inside Diameter of Clamp Jaws
2.54 cm (1 in.)